Discover the future of laboratory technology: Our fully automated system PA-3600 for immunohistochemistry and in situ hybridization combines maximum efficiency with reproducible quality.
Designed for high-demand applications in pathology and molecular diagnostics, this solution enables seamless, standardized processing of your samples – from antibody detection – to gene localization. Fully IVD-R compliant portfolio of reagents and ready to use optimized antibodies.
Whether for routine workflows or complex research projects, this instrument elevates your lab operations with intuitive controls, flexible protocol design, and reliable reagent management.
Supports baking, dewaxing, antigen retrieval, staining and counterstaining.
Compatible with both immunohistochemistry (IHC) and Compatible with both immunohistochemistry (IHC) and in situ hybridization (ISH) protocols.
Slides can be loaded at any time during operation without interrupting ongoing processes.
Scans 36 slides in under 30 seconds using QR code identification.
Each of the 36 staining modules has its own heating system.
Temperature range: ambient to 102 °C with fluctuation <0.5 °C for precise and consistent results.
RFID chip technology scans 40 reagent bottles in under 10 seconds.
Conical reagent bottle design minimizes waste and reagent residue.
Real-time liquid buffer monitoring alerts users to refill buffer or dispose of waste.
The reagents on the slide moves in a wave-like motion to increase antibody-antigen reaction, shorten incubation time and ensure thorough cleaning.
Solid cover diaphragms allows for high temperature HIER without reagent evaporation.
Up to 5 instruments can be controlled simultaneously, saving lab space and increasing throughput.
Our immunohistochemical detection systems combine simplicity, precision, and high sensitivity to deliver reliable and reproducible results. Designed for efficient workflows and consistent staining quality, they enable clear visualization of target antigens while minimizing background and hands-on time.
Our IVDR-compliant antibodies, specifically designed for use with the PA-3600, enable reliable in vitro diagnosis of malignant tumors. They support the precise detection of tumor-specific markers and deliver standardized, reproducible results for modern oncological diagnostics.
Process up to 36 different reaction staining protocols simultaneously. Ideal for both routine diagnostics and complex research applications.
Precise temperature and reagent control ensure uniform staining quality. Standardized protocols faciliate relilable comparison across experiments.
Automated scanning and loading reduce manual labor and error.
Offline operation capability ensures continuity even during software disconnection. Continous random access and short running time < 3,5 h (36 Slides)
Minimal reagent usage per slide.
Efficient recource use reduces operating costs and supports a more enviromentally friendly application.
Instrument type |
PA-3600 |
|---|---|
Applications |
IHC, ICC, FISH, CISH, Multiplex |
Slide capacity |
36 (3 x 12) |
Reagents |
40 (4 x 10) |
Reagent pipetting volume |
130μL |
Reagent bottle volume |
7mL primary antibodies / 15mL reagents |
Running time |
<3.5h (36 Slides) |
Waste management |
Separation of hazardous substances and non-critical waste liquids |
Control & Network |
Each instrument is equipped with its own computer. Up to 5 instruments can be connected to a network, allowing easy operation.. |
LIS compatability |
Compatible with HL7 interface standard. |
Waste liquid volumes |
Non-hazardous: <1.8L / Hazardous: <0.2L |
Dimensions |
1050mm (W) / 690mm (D) / 736mm (H) |
Weight |
135kg |
Temperature range |
20–102°C |